Corporate endpoints
Office laptops and desktops. General productivity and internet access.
Flat networks are how an office phishing email reaches a chromatography data system. We design segmented laboratory networks with deny-by-default policies between zones, built around what your instruments and applications actually require.

A laboratory that shares one flat network between reception PCs, guest Wi-Fi, CCTV and a GC-MS workstation has no meaningful boundary to defend. Segmentation gives each class of system its own zone, and makes every crossing between zones an explicit, documented decision rather than an accident of cabling.
The design below is a typical starting structure. The final scheme depends on your instrument map, application dependencies and what the instrument vendors support.
Office laptops and desktops. General productivity and internet access.
Laboratory staff workstations that handle regulated records but are not instrument-attached.
Instrument workstations and embedded controllers. The most tightly restricted zone.
CDS, LIMS, file, application and database services.
IP telephony, kept off the data zones.
Visitor access with no route to any internal zone.
Cameras, recorders and connected building devices, which are rarely patched and should never be adjacent to scientific data.
HVAC, environmental monitoring and facility controllers.
Backup infrastructure and administrative access, isolated so that ransomware reaching a user zone does not reach the backups.
Map every connected device, what it talks to, and which vendor supports it. Nothing is segmented before it is understood.
Zones, addressing, and the permitted flows between them, agreed with the laboratory and with instrument vendors where their software is affected.
Build and test the configuration out of hours, with a documented rollback for each change.
Move zone by zone, verifying instrument and application behaviour after each step rather than at the end.
Topology, VLAN register, firewall rule records and the justification for every inter-zone rule.
It can, if it is done without mapping dependencies first, which is why discovery comes before design. An HPLC workstation normally needs to reach its scientific application and database services, DNS and NTP, and specifically approved infrastructure — and nothing else. We establish that list with you and the vendor before any policy is enforced, and verify instrument behaviour after each cutover step.
Fewer than the list above if you are small, more if you have multiple buildings or a wide instrument estate. The useful question is not the count but whether any two systems with very different risk profiles share a zone. Instruments, backups and guest access should always be separated.
Usually. We assess what you have against the design before recommending replacement. Where existing equipment supports the required VLANs and policies, reusing it is normally the right call.
Pracoda provides technical infrastructure implementation and documentation support. Regulatory interpretation, GxP applicability decisions, validation approval, SOP approval, data review and final quality release remain with the client's authorized QA, system owner or qualified regulatory and validation specialists. Instrument software and scientific equipment support remain subject to the instrument vendor's terms.
Tell us roughly how many instruments and users you have and what the current network looks like. We will tell you what we would separate first.