ALCOA+ is the set of data integrity principles regulators use to judge whether laboratory data can be trusted. The original ALCOA acronym — attributable, legible, contemporaneous, original, accurate — came from good clinical practice, and the "+" extends it with complete, consistent, enduring and available. Together the nine principles appear across MHRA, FDA, WHO and PIC/S data integrity guidance.
For a QA professional, ALCOA+ is a framework for assessing records. For an IT team it is something more concrete: nine requirements that each map onto specific, checkable system configuration. Almost every data integrity finding in a laboratory traces back to a technical control that was absent, switched off, or available to the wrong people.
This guide takes each principle in turn, states what it means in practice, gives the technical controls that deliver it, and names the failure that regulators see most often.
Key takeaways
- Each ALCOA+ principle maps to concrete system configuration, not just procedure.
- Attributable fails first, and it almost always fails because of shared logins.
- "Original" means raw data and its metadata, not a printed summary of results.
- "Enduring" and "available" are storage, archival and migration problems that only show up years later.
- Technical controls beat procedural ones: a control a user cannot bypass needs no supervision.
Why ALCOA+ matters to IT
Data integrity findings are rarely about dishonesty. Far more often they describe a system that made the wrong thing easy: an instrument workstation where anyone could delete a folder, a shared account that made attribution impossible, a clock an analyst could change, a backup that had been failing silently for months.
That is why ALCOA+ is an infrastructure topic. Procedures tell people what to do; technical controls determine what they are able to do. A control that cannot be bypassed is worth more than an SOP requiring that it not be bypassed, and it costs less to maintain because it does not depend on anybody remembering.
Attributable
Every record and every change must be traceable to the person who made it, and to when.
Technical controls: unique named accounts on every system holding regulated data; no generic or shared logins; accounts never reassigned between people; a leaver process that disables rather than deletes accounts, so historical entries keep their meaning; named administrator accounts rather than a shared built-in administrator; audit trails that capture the user identity for each action.
The usual failure: a shared instrument account — LabUser, HPLC2, or the vendor engineer's installation account still in daily use. Once two people share credentials, no record on that system is attributable, and no amount of procedure recovers it. Fixing this is the highest-value data integrity work most labs can do, and it is what identity and device management addresses.
Legible
Records must be readable and permanent, and remain so for as long as they are retained.
Technical controls: store data in formats that can still be opened, not in undocumented proprietary blobs with no export path; keep the application version that can read the data available, or migrate to a format that outlives it; ensure exports render completely rather than truncating long fields; prevent storage on media that degrades or on individual workstations that get wiped at refresh time.
The usual failure: data acquired on a system that was decommissioned, in a format nothing current can open. The data physically exists on a backup tape; it is no longer legible in any practical sense.
Contemporaneous
Records must be made at the time the activity happened.
Technical controls: system clocks synchronised to a reliable internal time source; the user right to change system time removed; time changes logged; consistent time zone handling across the estate; audit trails time-stamped from the system clock rather than from user input; session timeouts so an unattended workstation cannot be used later under someone else's session.
The usual failure: an analyst able to change the clock on a standalone instrument PC. That single capability undermines every timestamp on the system, which is why time synchronisation appears in Part 11 technical controls despite seeming trivially administrative.
Original
The original record — or a verified true copy — must be preserved. In laboratory terms this means raw data with its metadata, not a summary.
Technical controls: raw data written to controlled, backed-up storage rather than a local disk; deletion of raw data prevented or restricted and logged; no overwrite of acquisition files; where a vendor requires local acquisition, a monitored and verified transfer to central storage; true copy processes that are documented and verified rather than a drag-and-drop by whoever was free.
The usual failure: treating the PDF report as the record. For chromatography and similar dynamic data, meaning depends on the integration, method and processing parameters — all of which a flattened report discards. Our guide to laboratory data infrastructure covers where raw data should actually live.
Accurate
Records must be correct, and changes must be visible and justified.
Technical controls: audit trails enabled and preserving the original value alongside the new one, with who and when; reason-for-change captured where required; audit trail not disableable by data-generating roles; calculations performed by validated software rather than uncontrolled spreadsheets; where spreadsheets are used for GxP calculations, they are locked, version-controlled and validated; review workflows enforced in the application rather than relied upon socially.
The usual failure: an uncontrolled spreadsheet in the calculation path, with live formulas any user can alter and no audit trail at all. It is the most common accuracy gap in otherwise well-run labs.
Complete
All data must be retained, including repeat analyses, failed runs, aborted injections and the audit trail — not only the results that were reported.
Technical controls: deletion of any acquired run prevented or restricted and logged; no capability to acquire data outside the controlled system; audit trail review that looks for deleted, aborted or repeated runs; system configured so that a test injection cannot be performed invisibly; storage sized so nobody is ever tempted to delete old data to make room.
The usual failure: selective reporting, whether deliberate or casual — reprocessing until a result passes and retaining only the final version. This is the pattern regulators look for hardest, and the technical defence is simply to make deletion impossible and repeats visible.
Consistent
Records must be in a consistent sequence, with date and time stamps that make sense in relation to each other.
Technical controls: a single synchronised time source across every system so sequences reconcile; consistent naming and numbering conventions enforced by the application; the same standard build and configuration across comparable workstations; consistent data paths rather than per-analyst folder habits.
The usual failure: two systems whose clocks differ by minutes, producing records where a sample appears to have been analysed before it was received.
Enduring
Records must survive for their full retention period, on media and in formats that last.
Technical controls: data on resilient storage with redundancy; backup with a tested restore and a dated record of that test; at least one copy off-site or immutable; archive strategy with a planned migration path as applications, operating systems and media age; retention periods configured in the system rather than left to memory; media refresh planned before it fails.
The usual failure: backups that run but have never been restored, so nobody knows whether the data endures or merely appears to. A restore test with a date and a name on it is the only evidence that counts — see laboratory backup and disaster recovery.
Available
Records must be retrievable for review or inspection throughout the retention period.
Technical controls: an inventory of what data exists and where; indexed, searchable archives rather than unlabelled tape; export of complete records including metadata and audit trail, demonstrated rather than assumed; access for reviewers and auditors that does not require handing out administrator rights; documented retrieval procedure with a realistic time to produce a record.
The usual failure: data that exists somewhere on archive media but cannot be located or read within an inspection timeframe. Available in principle, unavailable in practice.
The controls ranked by value
If a lab cannot do everything at once, the order matters. These are ranked by risk reduction per unit of effort.
| Priority | Control | Principles addressed | Effort |
|---|---|---|---|
| 1 | Unique accounts, shared logins removed | Attributable | Low to medium |
| 2 | Backup with tested, documented restore | Enduring, available | Low |
| 3 | Time synchronisation, clock change blocked | Contemporaneous, consistent | Low |
| 4 | Audit trails enabled and protected | Accurate, complete, attributable | Low to medium |
| 5 | Deletion of raw data prevented | Original, complete | Medium |
| 6 | Data moved off local disks to controlled storage | Original, enduring | Medium |
| 7 | Administrator rights separated from analysts | Accurate, complete | Medium |
| 8 | Spreadsheets removed from calculation paths | Accurate | Medium to high |
| 9 | Audit trail review process operating | Accurate, complete | Medium, ongoing |
| 10 | Archive and migration strategy | Legible, enduring, available | High |
The first four are cheap, fast and address six of the nine principles. Labs that start at item eight because it feels more serious tend to still have shared logins a year later.
Running a gap assessment
A technical data integrity gap assessment works system by system. For each one, record the systems inventory entry, whether regulated records are held, how users authenticate, whether accounts are unique, who holds administrator rights, whether the audit trail is on and who could switch it off, whether the clock is synchronised and protected, where raw data is written, who can delete it, whether backup runs and when a restore was last proved, and what the retention and archive position is.
Produce the output as a register with a risk rating and an owner per gap, not as prose. Remediate technical controls first — they are faster, cheaper and do not rely on behaviour — then procedures, then system replacement where a legacy platform simply cannot meet a principle. Where a gap cannot be closed, document the assessment, the compensating controls and the dated plan; that position is defensible, whereas an unexamined gap is a finding waiting to happen.
Frequently asked questions
What does the "+" in ALCOA+ add?
Four principles beyond the original five: complete, consistent, enduring and available. They extend the focus from the individual record to the whole data set over its full life, which is why they translate so directly into storage, backup and archival design.
Is ALCOA+ a regulation?
No. It is a framework that appears across regulatory data integrity guidance, including MHRA's GxP data integrity guidance, FDA data integrity guidance, WHO and PIC/S documents. The binding requirements come from GMP, GCP, GLP and rules such as 21 CFR Part 11; ALCOA+ is the lens used to assess them.
Which ALCOA+ principle fails most often?
Attributable, through shared logins on instrument workstations. It is also the most consequential, because once attribution is lost every other control on that system is weakened — the audit trail records that "LabUser" made a change, which tells an investigator nothing.
Are spreadsheets acceptable in a GxP laboratory?
They can be, but a spreadsheet performing a GxP calculation is a computerised system: it needs validation, version control, protection of formulas, access control and some means of audit. An unlocked spreadsheet with live formulas in the calculation path is a recognised accuracy and integrity gap.
Can an IT provider deliver data integrity compliance?
An IT provider can deliver the technical controls and the evidence that they work. Whether the resulting controls are adequate for your processes, and whether procedures and review are sound, is a QA judgement that remains with the laboratory.
Want to know which ALCOA+ principles your systems would fail today? Our data integrity services run a technical gap assessment across your instruments and systems and produce a prioritised remediation register, alongside our wider lab compliance frameworks work. Tell us what you are running and we will tell you what we would look at first.
Want an expert to look at your lab IT?
Tell us how your laboratory is set up. A senior lab IT consultant will reply within one working day with honest, practical advice. No obligation and no sales script.
- Reply within one working day
- Advice even if we are not the right fit
- We sign an NDA before any detail is shared